Medical Air Compressors and Vacuum Replacement Equipment/System
DEPARTMENT OF VETERANS AFFAIRS
The Grand Junction VA Health Care System requires the replacement of medical air compressor and vacuum equipment/systems for their main VA medical center. The scope of work includes providing and installing two medical air compressor systems and two medical vacuum systems. Each compressor system requires a capacity of 33 SCFM at 50 PSIG, 15 total horsepower, a 120-gallon vertical receiver tank, and a desiccant air dryer. The vacuum systems require a working capacity of 52 SCFM @ 19" Hg, a maximum total horsepower of 15 (7.5HP x 2), a 120-gallon vertical receiver tank, and a Total Oxidize Monitoring and Mitigation System. Installation includes removal of existing equipment, connection to existing piping and electrical systems, configuration, programming, testing, and verification of alarms. All work must comply with current standards and manufacturer requirements, including NFPA 99 and ASSE 6010 certifications for installers. The equipment must be suitable for operation at an elevation of 4,600 ft. A minimum 1-year warranty on workmanship is required.